15-4 z.one ultra and z.one ultra sp OPERATOR MANUAL
In the onscreen menu under the Preset tab, select one of the factory default
settings (General, etc.) to determine possible maladjustment of user presets.
Ensure an adequate supply of acoustic coupling gel and good patient/
transducer contact.
Switch to a different transducer to determine if the problem is
transducer-specific.
No (or poor quality) Color mode image or PW Doppler
Make sure the brightness and/or contrast settings for the LCD Display have
not been altered from proper settings. For information about LCD Display
settings, see
page 2-6.
NOTE: To ensure the quality of the grayscale resolution of the LCD Display, use
the grayscale test pattern (via the Diagnostic menu) available on the
SmartCart/
SmartCart sp
. Ensure that all shades can be visualized, making small
adjustments to brightness and contrast from factory defaults, as required, to see
all intensities.
Make adjustments to the various Doppler controls (Gain, Filter, Scale) to
attempt to resolve problem.
Ensure that the scanning angle between the transducer face and the direction
of blood flow is optimized.
In the onscreen menu under the Preset tab, select one of the factory default
settings (General, etc.) to determine possible maladjustment of user presets.
Ensure there is an adequate supply of acoustic coupling gel and good patient/
transducer contact.
Switch to a different transducer to determine if the problem is
transducer-specific.
No (or poor quality) CW Doppler
If using the CW Aux transducer:
Disconnect the CW Aux connector from the front of the SmartCart and then
reconnect to ensure electrical continuity.
Try using a different CW Aux transducer.
Try using a P4-1c transducer to perform CW mode operation.
If using the P4-1c transducer:
Try using the CW Aux transducer.
Unplug the P4-1c connector from the MTP and reconnect to a different port on
the MTP.
Try operating the P4-1c transducer in PW Doppler mode.
If system still fails to operate CW Doppler, contact ZONARE Technical
Support.
No ECG operation
Activate ECG operation by pressing the ECG key on the SmartCart QWERTY
keyboard.