9-48 z.one ultra and z.one ultra sp OPERATOR MANUAL
Enter, Enter/Select, or Select. Additionally, Autotrace Calc results can only be
entered from a frozen strip display (not 'live' or Cine Play).
• While Auto-Dop Trace is active, the Doppler envelope will be outlined in
green. When you freeze the strip, Automeasure numbers are displayed at the
bottom right side of the image.
• When the image is frozen, vertical BLUE line graphics delimiters will appear
on the PW strip. A small yellow arrow will appear at the top to mark the peak
systolic velocity. When you have multiple cardiac cycles displayed, a small
yellow arrow will also be displayed at the bottom to mark the peak end
diastolic.
• The placement of the delimiters is determined by how many heart cycles were
chosen in the N Cycles selection of the Auto Params menu. The heart cycle
selected for active evaluation can be changed by scrolling the Trackball to the
left or right.
• You can manually adjust the position of the delimiters by pressing the Set
button. The first press will allow for adjustment of the Left delimiter; the
second press will allowing adjustment of the Right delimiter. The calculated
Automeasure values will be adjusted accordingly.
• Auto-Dop Trace is different from conventional autotrace functions in that the
Auto-Dop Trace process is applied to the spectrum prior to any
post-processing functions (such as gain, dynamic range, edge) being applied.
• Manual adjustments to any of these post-processing functions will not affect
the Auto-Dop Trace.
NOTE: Auto-Dop Trace is intended to serve as an adjunct to the diagnostic
process in evaluating blood flow during PW Doppler examinations. When using
the Auto-Dop Trace feature, please evaluate the results to verify that you are in
agreement before committing the values to the Calc Report Package.