Quatech Audio/Video Capture USB 2.0 Especificações Página 252

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 251
9
9-50 z.one ultra and z.one ultra sp OPERATOR MANUAL
4. On next menu are three selections: Vessel Diam, PW Gate Size, and Start
Trace.
5. Highlight Vessel Diam. Then use Trackball or Set button to position horizontal
lines on inner diameter of vessel. Vessel diameter represents diameter of flow
area that blood passes through.
6. Next, highlight PW Gate Size. Then use Trackball or Set button to position PW
Gate within vessel diameter to encompass entire flow.
7. Next, select Start Trace and press Set.
8. The following generic numbers will be displayed at the bottom of the image:
Diameter, TAMN, PSV, Volume Flow (ml/min).
Auto Doppler
Volume Flow
Grafts/Shunts
(Vasc Calc)
Volume Flow Grafts/Shunts (Vasc Calc) is available only when using the Vasc/
Arterial-UE and Vasc/Arterial-LE presets.
To Measure Volume Flow Grafts/Shunts (Vasc Calc)
1. Press Exam Type button. Select Vascular exam softkey, then select
Arterial-UE or Arterial-LE softkey.
2. Press Function/Mode key assigned to Autotrace (see “To Activate Auto-Dop
Trace on page 7-51).
3. Press Calc button. Calc menu displays. Select Graft Vol Flow and press Set.
Figure 9-26. Calc Menu and Graft Vol Flow Submenu
4. Adjust Vessel Diam and PW Gate Size as needed (see “To Measure Volume
Flow” on page 9-49).
5. Select one of the Native, Anast, or Graft options and press Set.
6. The following measurements will display at the bottom of the image:
Diameter, TAMN, PSV, Volume Flow (ml/min).
7. To redisplay the Graft Vol Flow submenu, press Calc button.
Vista de página 251
1 2 ... 247 248 249 250 251 252 253 254 255 256 257 ... 403 404

Comentários a estes Manuais

Sem comentários