Quatech Audio/Video Capture USB 2.0 Especificações Página 363

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 362
14
z.one ultra and z.one ultra sp OPERATOR MANUAL 14-21
WARNING: Bent, broken, or missing pins on the transducer connector may cause
poor image quality, including possible mirror image artifact. Be sure to check pins
before connecting transducer to the
ZONARE ultrasound system. If pins are bent,
broken, or missing, do not use the transducer and call
ZONARE Technical
Support.
Figure 14-30. Bent, Broken, or Missing Pins on Transducer
Inspecting
Transducers
Inspect transducers at least weekly for the following problems.
Inspect each transducer for signs of wear or damage:
Cracks in case or transducer face
Cuts or gouges on any part of the transducer, including transducer face,
case, cable, and connector
Buckling or bulging of the lens material on the transducer face
Damage to the transducer connector, including bent, broken, or missing
pins
WARNING: Immediately replace a transducer that exhibits any of the above
damage symptoms.
Verifying
Imaging
Performance
Verify transducers for proper imaging performance at least weekly.
Scan yourself or a tissue-equivalent phantom with each transducer.
Verify uniform image quality. Look for signs of drop-out (dark columns in the
image area).
Cleaning and Disinfecting Transducers
Refer to page 6-14 for cleaning and disinfecting procedures for the P8-3 TEE
transducer.
Refer to “ZONARE Transducers Cleaning and Disinfection” (Q00066) for cleaning
and disinfecting procedures for all other transducers.
Bent pins
Missing/
damaged
pin
Vista de página 362
1 2 ... 358 359 360 361 362 363 364 365 366 367 368 ... 403 404

Comentários a estes Manuais

Sem comentários