7
z.one ultra and z.one ultra sp OPERATOR MANUAL 7-43
Pulsed Wave (PW) Doppler Imaging
Pulse Wave (PW) Doppler is used to assess blood flow velocity.
Figure 7-19. Pulsed Wave (PW) Doppler Image
NOTE: See Figure 10-16 on page 10-14 for a description of the Doppler Data
Field. See also Chapter 10 Echocardiography Option for information about PW
and cardiac imaging.
Simul: (Dual)
Scan Engine
Press Simul softkey to toggle On/Off
Simultaneous Dual imaging update
modality.
NOTE: When Simul Dual is active, the two
different images, potentially using different
modalities (i.e. Color Doppler & Power
Doppler) will be dynamically updated,
simultaneously.
NOTE: When this function is deselected
(standard Dual mode) one image will be
static, while other image is dynamically
active.
Highlight Simul: Dual. Press Select to
toggle On/Off this function.
NOTE: When Simul Dual is active, the two
different images, potentially using different
modalities (i.e., Color Doppler & Power
Doppler) will be dynamically updated,
simultaneously. When this function is
deselected (standard Dual mode) one image
will be static, while other image is
dynamically active.
Table 7-3. Color Doppler and Power Doppler Controls (Continued)
CD/PD Control SmartCart/SmartCart sp Scan Engine
Indicates that this control is also available in retrospective processing
Doppler Data
Field
Vista de página 146